U S E R ’ S G U I D EAUTOTUNE SERIESHIGH INTENSITY ULTRASONIC PROCESSORWITH TEMPERATURE CONTROLLERMICROPROCESSOR CONTROLLED (SERIAL NO. "Y"
7No DESCRIPTIONOrderNumber1 Converter Model CV33 CV000332 Four element coupler 630-05583 Stepped top(s) 1/8” (3mm) 630-05354 Booster BHNVCGD5 Probe ½”
8USING THE ULTRASONIC PROCESSORThe speed control on an automobile, can, to a certain extent, be compared to anUltrasonic Processor. The speed control
Press the I key. The screen will display the power rating and the frequency of theUltrasonic Processor, and the following control parameters.AMPLITUDE
1. Using the fittings on top of the converter, circulate dry compressed air through theconverter to cool the converter.2. Immerse the probe into the l
3. Depress the START key.4. Using the▲ or ▼ keys, increase or decrease the amplitude as required.TIMER: In the pulsed mode the processing time will be
PULSER: By inhibiting heat build-up in the sample, the pulse function enables safetreatment of temperature sensitive samples at high intensity. In add
TEMPERATURE: The temperature function prevents overheating of the sample bycontinuously monitoring the sample temperature, and terminating the ultraso
14REVIEW: The REVIEW function provides a “window” on the process by displayingvarious operating parameters without process interruption. Pressing theE
ENERGY SETPOINT: The energy setpoint continuously monitors the amount ofenergy in Joules (watts-seconds), that is being delivered to the probe, and te
RECALL: The recall function can retrieve from memory any of 9 stored controlparameters for verification or usage. To retrieve any parameters press the
WARRANTYYour Ultrasonic Processor is warranted for a period of three years from thedate of shipment against defects in material and workmanship under
SECTION III – SERVICE INFORMATIONYour Ultrasonic Processor was designed to provide you with years of safe anddependable service. Nevertheless, because
18RETURN OF EQUIPMENTIt is suggested that if a unit is in need of repair, it should be sent back to the factory.In order to receive prompt service; al
19SONICS SAFETY CERTIFICATION FORMItems being returned:Please check only one item below:___ The equipment was never used or exposed to any radiologica
20SECTION IV -OPERATING SUGGESTIONS AND TECHNIQUESDISRUPTING CELLSThe disruption of cells is an important method in the field of proteomics and in the
21Single-cell organisms (micro-organisms) consist of a semipermeable, tough, rigidouter cell wall surrounding the protoplasmic membrane and cytoplasm.
22The ability to control the amplitude at the probe tip is a prerequisite for processoptimization. And because each application requires its own set o
23the sample remains trapped in intact cells and, therefore, is unavailable forsubsequent purification. For most samples, thorough disruption can be m
24Detergent cell lysis is sometimes used in conjunction with ultrasonic processing tofacilitate disruption. Detergents break the lipid barrier surroun
25shown to effectively release recombinant proteins located in the cytoplasm of bacteriaand is recommended when lysing mammalian cells. However, this
26When processing a sample with ultrasonics, always immerse the probe deep enoughbelow the surface of the sample to inhibit aerosoling or foaming, foa
IMPORTANT SAFEGUARDSREAD BEFORE INSTALLING ORUSING THE EQUIPMENTYour Ultrasonic Processor has been designed with safety in mind. However, no designcan
27Before each application, place the tip in 100% ethanol and energize the power supplyfor a few seconds to remove any residual substances. If still co
28Fractional protein release, Rp, is calculated using the following equation andmultiplying the result by 100:Rp = Cf – CbCt – CbCf = Free proteinCt =
1SECTION I – INSTALLATIONINSPECTIONPrior to installing the Ultrasonic Processor, perform a visual inspection to detect anyevidence of damage which mig
2SECTION II – OPERATIONPRINCIPLES OF ULTRASONIC DISRUPTIONThe ultrasonic power supply converts 50/60 Hz line voltage to high frequencyelectrical energ
Displays prompts and the following control parameters:• Amplitude selected• Output power delivered to the probe in watts, and as percentage of the tot
FUNCTIONS OF KEYS, CONTROLS, INDICATORS, AND CONNECTORS (cont.)REAR PANEL9-pin D-sub connector Connects to external actuation device, and enable power
PREPARATION FOR USE1. Plug the electrical line cord into the electrical outlet. If the unit is already on, switchthe unit off by depressing the O key.